Sensor Connector Cell con‐
stant
Cell con‐
stant (1/cm)
T-correction
element
Max. temp.
(°C)
Measuring
range κ min
(Unit)
Measuring
range κ
max
(Unit)
LM1-TA Fixed cable
0.34 mm
2
,
5 m,
shielded
1.00 - 70 0.1 mS/cm 20 mS/cm
LMP1 DIN 4-pin 1.00 Pt100 70 0.1 mS/cm 20 mS/cm
LMP1-HT DIN 4-pin 1.00 Pt100 120 0.1 mS/cm 20 mS/cm
LMP1-TA Fixed cable
0.34 mm
2
,
5 m,
shielded
1.00 Pt100 70 0.1 mS/cm 20 mS/cm
CK1 DIN 4-pin 1.00 - 150 0.01 mS/cm 20 mS/cm
CKPt1 DIN 4-pin 1.00 Pt100 150 0.01 mS/cm 20 mS/cm
7.4 Priming to bleeding
The pump is working at 100%
performance
Note any installation work in your sur‐
roundings, as feed chemical can
uncontrollably escape into the environ‐
ment in the event of open pipes etc.
Pump 1
A1068
Function Decrease value
Max. stroke rate
Assignment Channel 1
Fig. 29: [Prime with <OK>] e.g. to bleed a pump
If you select the function
[Prime with <OK>]
when the pumps are connected and operable,
the pumps continue to operate at 100% power
for as long as you press and hold down the
key.
You can use this function, for example, to
transport the feed chemical to the pump,
thereby bleeding the metering line.
Assembly and installation
65