EasyManuals Logo

ProMinent DULCOMETER diaLog DACb User Manual

ProMinent DULCOMETER diaLog DACb
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #177 background imageLoading...
Page #177 background image
Parameter Measuring range/Measured value
Specific conductivity: 0.001 µS/cm ... 200 mS/cm
Specific electrical resistance: 5 Ωcm ... 1000 MΩcm
TOS (total dissolved solids): 0 ... 9999 ppm (mg/l)
SAL (salinity): 0.0 ... 70.0 ‰ (g/kg)
19.2 Technical data
Tab. 26: Technical data
Description Technical data
pH resolution: 0.01
ORP voltage: 1 mV
Temperature: 0.1 °C
Amperometric analysis (chlorine etc.): 0.001/0.01 ppm, 0.01% vol., 0.1% vol.
Precision: 0.3% based on the full-scale reading
pH/ORP measurement input: Input resistance > 0.5 x 1012 Ω
Correction variable: Temperature via Pt 100/Pt 1000
Temperature correction range: 0 … 100 °C
pH correction range for chlorine: 6.5 … 8.5
Interference variable: Flow via mA or frequency
Control action: P/PID control
Control: 2 bidirectional controllers or 1 bidirectional controller and
1 monodirectional controller
mA output signal: 2 x 0/4 ... 20 mA galvanically isolated, max. load 450 Ω,
range and assignment (measured, correction, control
variable) can be set
Control output: 2 x 2 pulse frequency outputs for control of metering
pumps
2 relays (limit value, 3-point step or pulse length control)
Measuring range and technical data
177

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ProMinent DULCOMETER diaLog DACb and is the answer not in the manual?

ProMinent DULCOMETER diaLog DACb Specifications

General IconGeneral
BrandProMinent
ModelDULCOMETER diaLog DACb
CategoryController
LanguageEnglish

Related product manuals