Selection of the connected sensor
All of the sensor-dependent settings are reset to the [DEFAULT] values when changing the
connected sensor.
Sensor Connector Cell con‐
stant
Cell con‐
stant (1/cm)
T-correction
element
Max. temp.
(°C)
Measuring
range κ min
(Unit)
Measuring
range κ
max
(Unit)
LFTK1FE3
m
Fixed cable
0.25 mm
2
, 3
m, shielded
1.00 Pt1000 80 0.01 mS/cm 20 mS/cm
LFTK1FE5
m
Fixed cable
0.25 mm
2
,
5 m,
shielded
1.00 Pt1000 80 0.01 mS/cm 20 mS/cm
LFTK1-DE DIN 4-pin 1.00 Pt1000 80 0.01 mS/cm 20 mS/cm
LFTK1-1/2 DIN 4-pin 1.00 Pt1000 80 0.01 mS/cm 20 mS/cm
LF1-DE DIN 4-pin 1.00 - 80 0.01 mS/cm 20 mS/cm
LFT1-DE DIN 4-pin 1.00 Pt100 80 0.01 mS/cm 20 mS/cm
LFT1-1/2 DIN 4-pin 1.00 Pt100 80 0.01 mS/cm 20 mS/cm
LMP01 DIN 4-pin 0.10 Pt100 70 0.1 uS/cm 500 uS/cm
LMP01-HT DIN 4-pin 0.10 Pt100 120 0.1 uS/cm 500 uS/cm
LMP01-TA Fixed cable
0.34mm
2
,
5 m,
shielded
0.10 Pt100 70 0.1 uS/cm 500 uS/cm
LMP001 DIN 4-pin 0.01 Pt100 70 0.01 uS/cm 50 uS/cm
LMP001-HT DIN 4-pin 0.01 Pt100 120 0.01 uS/cm 50 uS/cm
LM1 DIN 4-pin 1.00 - 70 0.1 mS/cm 20 mS/cm