EasyManuals Logo

ProMinent DULCOMETER diaLog DACb User Manual

ProMinent DULCOMETER diaLog DACb
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #64 background imageLoading...
Page #64 background image
Selection of the connected sensor
All of the sensor-dependent settings are reset to the [DEFAULT] values when changing the
connected sensor.
Sensor Connector Cell con‐
stant
Cell con‐
stant (1/cm)
T-correction
element
Max. temp.
(°C)
Measuring
range κ min
(Unit)
Measuring
range κ
max
(Unit)
LFTK1FE3
m
Fixed cable
0.25 mm
2
, 3
m, shielded
1.00 Pt1000 80 0.01 mS/cm 20 mS/cm
LFTK1FE5
m
Fixed cable
0.25 mm
2
,
5 m,
shielded
1.00 Pt1000 80 0.01 mS/cm 20 mS/cm
LFTK1-DE DIN 4-pin 1.00 Pt1000 80 0.01 mS/cm 20 mS/cm
LFTK1-1/2 DIN 4-pin 1.00 Pt1000 80 0.01 mS/cm 20 mS/cm
LF1-DE DIN 4-pin 1.00 - 80 0.01 mS/cm 20 mS/cm
LFT1-DE DIN 4-pin 1.00 Pt100 80 0.01 mS/cm 20 mS/cm
LFT1-1/2 DIN 4-pin 1.00 Pt100 80 0.01 mS/cm 20 mS/cm
LMP01 DIN 4-pin 0.10 Pt100 70 0.1 uS/cm 500 uS/cm
LMP01-HT DIN 4-pin 0.10 Pt100 120 0.1 uS/cm 500 uS/cm
LMP01-TA Fixed cable
0.34mm
2
,
5 m,
shielded
0.10 Pt100 70 0.1 uS/cm 500 uS/cm
LMP001 DIN 4-pin 0.01 Pt100 70 0.01 uS/cm 50 uS/cm
LMP001-HT DIN 4-pin 0.01 Pt100 120 0.01 uS/cm 50 uS/cm
LM1 DIN 4-pin 1.00 - 70 0.1 mS/cm 20 mS/cm
Assembly and installation
64

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ProMinent DULCOMETER diaLog DACb and is the answer not in the manual?

ProMinent DULCOMETER diaLog DACb Specifications

General IconGeneral
BrandProMinent
ModelDULCOMETER diaLog DACb
CategoryController
LanguageEnglish

Related product manuals