You may need a manual measuring
instrument for the temperature measured
variable. This manual instrument should
measure and display sufficiently accu‐
rately to guarantee successful calibration.
1.
Press in the continuous display.
2.
Then press
3. Follow the instructions in the con‐
troller display and perform calibra‐
tion
4.
Then press
5.
Then press to adjust the value or
press
to continue with calibration
6. Import the result of the calibration
into the controller memory by
pressing
ð
The controller shows the con‐
tinuous display again and oper‐
ates with the results of the cali‐
bration.
Incorrect calibration
An error message appears
if the result of the calibra‐
tion lies outside the speci‐
fied tolerance limits. In this
case the current calibration
is not carried over.
Check the prerequisites for
calibration and eliminate
the error. Then repeat cali‐
bration.
Calibration
100