Calibration specifica‐
tions of the sensor manu‐
facturer
When determining the cali‐
bration interval, also take
into consideration the oper‐
ating instructions for the
sensor, which may offer
additional and/or different
calibration intervals.
9.5.1
Selection of the calibration
process for the measured
variable O
2
To calibrate the controller there are three
available calibration processes:
n automatic
n DO value
n Zero point
Calibration process selection
1.
Continuous display ➨
CAL DO
A1049
Zero point
Slope
DO value
Zero point
automatic
Fig. 62: Calibration menu [DO]
ð
The calibration menu is dis‐
played.
2. Using the arrow keys select the
desired menu item. Press the
key
ð
You can now start the selected
calibration process.
Calibration
91

Table of Contents

Other manuals for ProMinent DULCOMETER

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ProMinent DULCOMETER and is the answer not in the manual?

ProMinent DULCOMETER Specifications

General IconGeneral
Protection ClassIP65
Measurement ParameterspH, Conductivity, Temperature
CommunicationModbus
Power Supply24 V DC or 100-240 V AC
Control OutputsRelay, Analog (0/4-20 mA)
Control FunctionsPID control
Input SignalsAnalog (0/4-20 mA), Digital
MountingWall-mounted, panel-mounted
Measurement Range pH0-14 pH
Measurement Range Conductivity0-2000 µS/cm

Related product manuals