19.4 Update
Continuous display ➨ ➨ or
[Setup]
➨ f
[Device setup]
➨ or
[Update]
➨
[Update]
A software update may be necessary if:
n New functions or new operating lan‐
guages are available and the software
is to be upgraded with them
n A software modification is necessary.
In this case you will be informed by
ProMinent or its plant engineer /
dealer about it.
A software update does not change the
current device settings.
The following are required for a software
update:
n A PC with internet access, so that you
can download the necessary software
n A PC with an SD card reader
n An SD memory card, maximum size
16 GB, for software transfer
You can download the latest software
from the link on the ProMinent homepage:
http://www.prominent.de/
desktopdefault.aspx/
tabid-12145/1485_read-67006/
, the
[Firmware DACa]
can be found under the
link Info/Downloads. Ø Create a directory
called Update on the memory card.
Under the above link you will find a ZIP
file containing 4 files:
n DACa_Lan.plf
n EXTa_up.mhx
n DACa_up.mhx
n info.txt
You can find out the latest software ver‐
sion in the file
‘info.txt’
. Copy all four files
into the
‘Update’
directory of the SD card.
An update is carried out in 3 steps:
n Update of the main module =
[Base board]
n Update of the expansion module =
[Ext board]
n Update of the language file =
[Language]
1. To do this access the following
menu in the controller and carry out
the following 3 steps:
ð
The controller reads in the
respective data, thereafter the
display goes out for approxi‐
mately 30 seconds and then
the controller reboots. The
steps must be carried out man‐
ually.
2.
Main module: ➨ ➨ or
[Setup]
➨
[Device setup]
➨ or
[Update]
➨ or
[Base board]
➨
ð
Updating starts
3.
Expansion module: ➨ ➨ or
[Setup]
➨
[Device setup]
➨
or
[Update]
➨ or
[Ext board]
➨
ð
Updating starts
4.
Language file: ➨ ➨ or
[Setup]
➨
[Device setup]
➨ or
[Update]
➨ or
[Language]
➨
Setting
[Device setup]
161