EasyManua.ls Logo

ProMinent DULCOMETER - Calibrating the Fluoride Sensor

ProMinent DULCOMETER
180 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9.3
Calibrating the Fluoride
Sensor
9.3.1
Selection of the calibration
process for fluoride
To calibrate the controller there are two
available calibration processes:
n 1 point
n 2 point
Calibration process selection
1.
Continuous display ➨
CAL F
-
A1037
-59.16 mV/dec
Slope
1 ppm = 185.0 mV 16:51:18
Single point calibration
11:11:11
11/11/2011
100 %
11/11/2011
Two point calibration
Fig. 50: Calibration menu [Fluoride]
ð
The calibration menu is dis‐
played.
2. Using the arrow keys select the
desired menu item. Press the
key
ð
You can now start the selected
calibration process.
9.3.2
2-point fluoride sensor cali‐
bration (CAL)
Correct sensor operation
Correct measuring and metering
is only possible if the sensor is
working perfectly
Observe the sensor operating
instructions
The carrying out of a 2-point cali‐
bration is strongly recommended
and is to be preferred to other
methods
For calibration the sensor must
be removed and refitted in the in-
line probe housing. To do this,
observe the operating instructions
of your in-line probe housing
Material required for calibration of fluoride
sensors:
n Two test containers with calibrating
solution
Calibration
80

Table of Contents

Other manuals for ProMinent DULCOMETER

Related product manuals