EasyManua.ls Logo

ProMinent DULCOMETER - Selecting the Calibration Process for Ph

ProMinent DULCOMETER
180 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAL pH
Asymmetry
Buffer 2:
A1019
Zero point
Slope
% Slope
Accept with <CAL>
Buffer 1:
Calibr.param. for 25 °C
Fig. 31: Display of the calibration result
CAL pH
Asymmetry in mV
Sensor quality
A1481
Slope in mV/pH
acceptable
good
good
55 56 57 58 59 60 61 62
-60 -30 0 30 60
Fig. 32: Displayed after the key has
been pressed
9.1.1
Selecting the Calibration
Process for pH
To calibrate the controller there are three
available calibration processes:
n 2-point
n Sample (1-point)
n Data input
Selecting the calibration process
1.
Continuous display ➨
ð
The Calibration menu is dis‐
played, you may need to select
[Channel 1]
or
[Channel 2]
,
depending on the measuring
channel on which the pH meas‐
urement is performed.
2.
Press
CAL pH
Buffer manufacturer
Calibration process
A1025
Off
Buffer detection
2 point
requirement
Buffer value 1
Buffer value 2
Buffer temperature
ProMinent
pH 7
pH 4
Fig. 33: Selecting the calibration process
ð
The menu for selecting the cali‐
bration process appears.
3. Use the arrow keys to select the
required menu item and press
ð
The input window appears and
you can make the necessary
settings for your process
4. Use the arrow keys to select the
calibration process and press
5.
Continue with
ð
You can now start your chosen
calibration process.
Calibration
65

Table of Contents

Other manuals for ProMinent DULCOMETER

Related product manuals