The following adjustable parameters are available when selecting the
[Measured value]
,
[Control variable]
and
[Correction value]
functions:
[Func‐
tion ]
Adjustable value Adjustable ranges or counter values
[Meas‐
ured
value]
[Control
variable]
[Correc‐
tion
value]
[Output range ]
0 ... 20 mA
Assignment to the required measuring range start
and end value.
4 ... 20 mA
Assignment to the required measuring range start
and end value.
[Error current] [Off]
23 mA
[0 mA]
- 100 % ... 100 %
[20 mA]
- 100 % ... 100 %
[Filtering] [high]
[medium]
[weak]
[Response with Pause
Hold]
[None]
The mA output changes with the measured value
[Fixed]
The mA output is set to a fixed mA output value,
which is always issued at
[Pause Hold]
[Hold]
Setting the
[mA outputs]
139