EasyManua.ls Logo

ProMinent DULCOMETER - Page 77

ProMinent DULCOMETER
180 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Measuring and control behaviour
of the controller during calibration
During calibration the actuating out‐
puts are deactivated. Exception: a
basic load or a manual control vari‐
able has been set. This remains
active. The measured value output
[standard signal output mA] is frozen,
corresponding to its settings in the
mA output menu.
When calibration/testing has been
completed successfully, all of the
error checks relating to the measured
value are restarted. The controller
saves all the determined data for zero
point and slope upon a successful
calibration.
Buffer used
Dispose of the used buffer solution.
For more information: refer to the
buffer solution safety data sheet.
You need one test container with a buffer
solution for calibration.
Continuous display ➨
CAL ORP
Pot. equalisation
Calibration process
A1027
No
0.0 mV
Offset
Last calibration
CAL setup
13:26:11
11/04/2013
continue with <CAL>
1 point
Fig. 44: 1-point calibration of ORP sensor
(CAL)
1.
Continue with
CAL ORP
A1028
Immerse sensor in buffer
continue with <CAL>
Fig. 45: 1-point calibration of ORP sensor
(CAL)
2. Carry out the instructions and then
press
ð
Calibration is running .
[Please wait!]
flashes.
Calibration
77

Table of Contents

Other manuals for ProMinent DULCOMETER

Related product manuals