EasyManua.ls Logo

ProMinent DULCOMETER - Page 97

ProMinent DULCOMETER
180 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1.
Press the key in the continuous
display.
2. Using the arrow keys select
[DO value]
3.
Then press
CAL DO
A1074
Air temperature
Water temp.
continue with <CAL>
Air pressure
Adjusting the concentration
higher than
Sea level
Relat. humidity
Salinity of the
Water
Fig. 68: DO value calibration for the meas‐
ured variable DO
4.
Then press to adjust the values
or press to continue with the cal‐
ibration
CAL DO
A1076
DPD value
Start with <CAL>
Sensor value 0.00 ppm
Sensor current
8.03 mA
Fig. 69: DO value calibration for the meas‐
ured variable DO
5.
Then press
CAL DO
A1077
1) Take sample
Change with <OK>
7.04 ppm
continue with <CAL>
2) Determine DPD value
Fig. 70: DO value calibration for the meas‐
ured variable DO
6. Take a water sample and determine
the DPD value with a suitable
measuring instrument.
7.
Then press to adjust the values
or press to continue with the cal‐
ibration
CAL DO
A1078
Calibration successful
Zero point
continue with <CAL>
Slope
Fig. 71: DO value calibration for the meas‐
ured variable DO
8. Import the result of the calibration
into the controller memory by
pressing the key
ð
The controller displays the con‐
tinuous display again and oper‐
ates with the results of the cali‐
bration.
Calibration
97

Table of Contents

Other manuals for ProMinent DULCOMETER

Related product manuals