ULTRA 4 INSTRUCTION MANUAL
58
To program the unit for Example 1 using the Quick Setup menu, proceed
as follows.
If required access Program Mode, by keying in the passcode 1997 and
press ENTER.
Using the ‘right’ arrow key, go to the Quick Setup menu and press ENTER.
And as prompted, by the questions, select the relevant option and press
ENTER.
QUESTION
OPTION
Level/Volume
1 = Level App.
No. of alarms
2 = 2 Alarms
Type alarm 1
1 = High
Alarm no.1
1 = Set to relay 1
Type alarm 2
2 = Low
Alarm no.2
2 = Set relay 2
Xducer (P101)
2 = dB6
Material (P102)
1 = Liquid
Measurement units (P104)
1 = Metres
Empty Level (P105)
3.5m
Span (P106)
2.8m
Programming is now complete, and the unit can be returned to the run
mode, press CANCEL until Run Mode? Is displayed on the LCD press
ENTER, and the Ultra 4 will return to the Run Mode.
Important Notice
If relay setpoints do not meet the exact requirements of the application,
they can be modified to suit by pressing ENTER when, “For More Options
Hit Enter”, is displayed, and entering new values to relay setpoints as
required. Alternatively, the relevant relay setpoint can be accessed either
by the main menu system or directly via parameter number and changed
as necessary.