For Zone 1 applications a transducer certified to Sira 02ATEX5104X is
used, and must be supplied via a 4000A breaking fuse, which is fitted as
standard to the Ultra 5.
For Zone 0 a transducer certified to Sira 02ATEX2103X is used, which
must be connected to the Ultra 5 controller via an external Zener barrier.
See transducer label for certification details.
Single Transducer mode is used to measure space, level, distance, volume
or flow and the transducer should be connected to Transducer 1 input
terminals.
Dual Transducer mode is used to measure Differential and Average. For
Differential, the upstream transducer should be connected to Transducer
1 input terminal and the downstream transducer to Transducer 2 input
terminal. In cases where the Ultra 5 is required to measure Average, then
transducers can be connected to either one.
Relay Outputs
The five relays can be programmed for a variety of alarms, pump control, or
other process functions. The relay contacts are all rated at 5A at 240V AC.
All connections should be such that the short circuit capacity of the circuits
to which they are connected, is limited by fuses rated so that they do not
exceed the relay rating.
Current Output
This is an isolated (floating) mA output (to 150 V), of 4 - 20mA or 0 -
20mA, and the load should not exceed 500
.
Current Input (Optional)
This feature is available as an option only. Please consult Pulsar for further
details. The current input is an isolated (floating) mA input (to 150 V), 4 -
20mA or 0 -20mA.
Temperature Input (Optional)
The external temperature sensor allows more localised compensation of the
measured distance due to changes in temperature.
There are two models, Type A and Type B as follows:
Type A
-25ºC to 50ºC
Type B
-25ºC to 125ºC