EasyManua.ls Logo

R&S FPC Series - CalculateN:Marker:function:ddemod:cferror[:Result]; CalculateN:Marker:function:ddemod:mdepth[:Result]; CalculateN:Marker:function:ddemod:mdeviation[:Result]

R&S FPC Series
419 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
FPC
315User Manual 1178.4130.02 ─ 07
Example: //Query carrier frequency drift
CALC:MARK:FUNC:DDEM:CFDR?
Usage: Query only
CALCulate<n>:MARKer:FUNCtion:DDEMod:CFERror[:RESult]?
This command queries the carrier frequency error of the measured signal.
Suffix:
<n>
.
irrelevant
Return values:
<Frequency> <numeric value>
Default unit: Hz
Example: //Query carrier frequency error
CALC:MARK:FUNC:DDEM:CFER?
Usage: Query only
CALCulate<n>:MARKer:FUNCtion:DDEMod:MDEPth[:RESult]?
This command queries the modulation depth of the measured signal.
Prerequisites for this command
Select ASK demodulation ([SENSe:]DDEMod).
Suffix:
<n>
.
irrelevant
Return values:
<ModIndex> <numeric value>
Default unit: PCT
Example: //Query modulation depth
CALC:MARK:FUNC:DDEM:MDEP?
Usage: Query only
CALCulate<n>:MARKer:FUNCtion:DDEMod:MDEViation[:RESult]?
This command queries the frequency deviation of the measured signal.
Prerequisites for this command
Select FSK demodulation ([SENSe:]DDEMod).
Suffix:
<n>
.
irrelevant
Return values:
<Frequency> <numeric value>
Default unit: Hz
Remote Commands of the Digital Demodulation

Table of Contents