The Receiver Application
R&S
®
FPC
143User Manual 1178.4130.02 ─ 07
4. Select the "Range / Ref Position" menu item to open the corresponding menu.
5. Select the "Ref Position <x>" menu item.
6. Move the position of the reference level on the bargraph scale to another position.
The range is from 0 to 10. "0" corresponds to the left side of the bargraph, "10" cor-
responds to right side of the bargraph.
You can also select the level range that the bargraph covers. In the default state,
the bargraph covers 100 values of the unit you have selected (e.g. 100 dBμV).
7. Select the "Range <x>" menu item.
8. Define the level range you need.
Tip: Auto range. When you select the "Auto Range" menu item, the R&S FPC
automatically adjusts the range of the bargraph based on the currently applied sig-
nal strength.
15.4 Bandwidth Configuration
Access: "BW"
Access: "Setup" > "Config Overview" > "Bandwidth" (selected settings)
Remote commands to configure the bandwidth:
Chapter 19.13.4, "Bandwidth", on page 299
The receiver application adds 6 dB resolution bandwidths to the 3 dB resolution band-
widths already available in other applications. The 6 dB bandwidths are special band-
widths that are necessary for measurements according to CISPR16.
The R&S FPC provides an automatic bandwidth selection for CISPR bandwidths.
When you select the bandwidth automatically, the R&S FPC selects an appropriate
CISPR bandwidth, depending on the receiver frequency.
Frequencies < 150 kHz: 200 Hz CISPR bandwidth
Frequencies from 150 kHz to 30 MHz: 9 kHz CISPR bandwidth
Frequencies from 30 MHz to 1 GHz: 120 kHz CISPR bandwidth
Frequencies > 1 GHz: 1 MHz CISPR bandwidth
Selecting the bandwidth
1. Press the "BW" key to open the bandwidth menu.
2. Select one of the menu items.
"RBW: Manual": Selects a 3 dB bandwidth.
"CISPR BW: Manual": Selects a 6 dB bandwidth.
"CISPR BW: Auto": Automatically selects a 6 dB bandwidth, depending on the
receiver frequency (see table above).
Bandwidth Configuration

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FPC Series and is the answer not in the manual?

R&S FPC Series Specifications

General IconGeneral
BrandR&S
ModelFPC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals