Remote Control
R&S
®
FPC
335User Manual 1178.4130.02 ─ 07
DISPlay<n>:VSWR:Y:SCALe:MINimum <BottomValue>
This command defines the top value of the vertical axis for the SWR measurement for-
mat.
Prerequisites for this command
Select reflection or DTF measurement (MEASurement<n>:MODE).
Select SWR format (MEASurement<n>:FORMat).
Suffix:
<n>
.
irrelevant
Parameters:
<BottomValue> <numeric value>
Range: 1.1 to 71
Example: //Define lower value of the y-axis in SWR
MEAS:MODE DTF
MEAS:FORM VSWR
DISP:VSWR:Y:SCAL:MIN 1
Manual operation: See "Range / Ref Position" on page 399
19.16.3 Frequency
Commands to configure the frequency described elsewhere.
Center frequency:
[SENSe:]FREQuency:CENTer
Span:
[SENSe:]FREQuency:SPAN
Start frequency:
[SENSe:]FREQuency:STARt
Stop frequency:
[SENSe:]FREQuency:STOP
Frequency mode (frequency / channel):
[SENSe:]FREQuency:INPut:MODE
Channel selection:
[SENSe:]CHANnel
[SENSe:]FREQuency:SETTings:COUPling:ENABle...........................................................335
[SENSe:]FREQuency:SPAN:AUTO..................................................................................336
[SENSe:]FREQuency:SETTings:COUPling:ENABle <State>
This command couples or decouples the frequency settings of the DTF measurement
to the other VNA measurements.
Remote Commands of the VNA Application