Remote Control
R&S
®
FPC
322User Manual 1178.4130.02 ─ 07
19.15.5 Sweep Control
Commands to control the measurement described elsewhere.
●
Measurement control:
INITiate[:IMMediate]
●
Measurement type (single / continuous):
INITiate:CONTinuous
19.15.6 Trigger
Commands to configure trigger described elsewhere.
●
Trigger slope:
TRIGger[:SEQuence]:SLOPe
TRIGger[:SEQuence]:LEVel:IQPower...............................................................................322
TRIGger[:SEQuence]:SOURce........................................................................................322
TRIGger[:SEQuence]:LEVel:IQPower <Level>
This command defines the trigger level of the I/Q power trigger.
Prerequisites for this command
●
Select I/Q power trigger (TRIGger[:SEQuence]:SOURce).
Parameters:
<Level> <numeric value>
*RST: -20
Default unit: dBm
Example: //Configure I/Q power trigger
TRIG:SOUR IQP
TRIG:LEV:IQP -10
Manual operation: See "Trigger" on page 394
TRIGger[:SEQuence]:SOURce <TriggerSource>
This command selects the trigger source.
Parameters:
<TriggerSource> IMMediate
No trigger is applied.
EXTernal
Select an external trigger.
IQPower
Select the I/Q power trigger.
Define the trigger level with TRIGger[:SEQuence]:LEVel:
IQPower.
*RST: IMMediate
Remote Commands of the Digital Demodulation