Remote Control
R&S
®
FPC
263User Manual 1178.4130.02 ─ 07
Manual operation: See "Continuous Measurement" on page 168
See "Single Measurement" on page 169
See "Continuous Sweep" on page 372
See "Single Sweep" on page 372
See "Continuous Scan" on page 382
See "Single Scan" on page 382
See "Continuous Measurement" on page 390
See "Single Measurement" on page 390
See "Continuous Measurement" on page 394
See "Single Measurement" on page 394
See "Continuous Sweep" on page 400
See "Single Sweep" on page 401
INITiate[:IMMediate]
The command initiates a new measurement sequence.
Effects of this command
●
In single sweep mode, with an average count > 0, the command restarts the mea-
surement.
●
When you have selected the maxhold, minhold or average trace mode the com-
mand resets the trace.
in single sweep mode, you can synchronize to the end of the measurement with *OPC,
*OPC? or *WAI. In continuous sweep mode, synchronization to the end of the mea-
surement is not possible. Thus, it is not recommended that you use continuous sweep
mode in remote control, as results like trace data or markers are only valid after a sin-
gle sweep end synchronization.
Example:
//Start a single sweep and wait to end of sweep
INIT:CONT OFF
INIT;*WAI
Usage: Event
Manual operation: See "Continuous Measurement" on page 168
See "Single Measurement" on page 169
See "Continuous Sweep" on page 372
See "Single Sweep" on page 372
See "Continuous Scan" on page 382
See "Single Scan" on page 382
See "Continuous Measurement" on page 390
See "Single Measurement" on page 390
See "Continuous Measurement" on page 394
See "Single Measurement" on page 394
See "Continuous Sweep" on page 400
See "Single Sweep" on page 401
Remote Commands of the Spectrum Application