Remote Control
R&S
®
FPC
262User Manual 1178.4130.02 ─ 07
Effects of this command
Turns off automatic selection of the sweep time ([SENSe:]BANDwidth[:
RESolution]:AUTO).
Parameters:
<SweepTime> <numeric value>
The available value range is specified in the datasheet.
*RST: - (AUTO is set to ON)
Default unit: s
Example: //Define sweep time
SWE:TIME 10s
Manual operation: See "Sweep Time" on page 365
See "Sweep Time: Manual" on page 372
See "Measurement Time" on page 382
[SENSe:]SWEep:TIME:AUTO <State>
This command couples and decouples the sweep time to the span and the resolution
and video bandwidths.
Effects of this command
Turn on automatic sweep time selection to couple the sweep time to the span and
the RBW.
When you turn off automatic selection of the sweep time, you can select one manually
with [SENSe:]SWEep:TIME.
Parameters:
<State> ON | OFF | 1 | 0
*RST: ON | 1
Example: //Turn on automatic selection of the sweep time
SWE:TIME:AUTO ON
Manual operation: See "Sweep Time: Auto" on page 372
INITiate:CONTinuous <State>
This command selects the sweep mode.
To actually start the measurement, use INITiate[:IMMediate].
Parameters:
<State> ON | 1
Select continuous sweep mode.
OFF | 0
Select single sweep mode.
Example: //Turn on single sweep mode.
INIT:CONT OFF