EasyManuals Logo

R&S NRP P Series User Manual

R&S NRP P Series
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #136 background imageLoading...
Page #136 background image
Remote control basics
R&S
®
NRPxxP
136User Manual 1179.5760.02 ─ 02
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
0
Sensor measuring
0
0
0
0
0
0
0
0
0
0
0
0
0
0
+
Figure 10-9: Operation measuring status register
Querying the register:
STATus:OPERation:MEASuring:CONDition?
STATus:OPERation:MEASuring[:SUMMary][:EVENt]?
Table 10-11: Used operation measuring status bits and their meaning
Bit no. Short description Bit is set if
1 Sensor measuring Power sensor is measuring.
10.2.5.3 Operation trigger status register
The CONDition register contains information whether a power sensor is waiting for a
trigger event. The EVENt register contains information whether the power sensor has
been waiting for a trigger event since the last query.