Remote control commands
R&S
®
NRPxxP
57User Manual 1179.5760.02 ─ 02
period to the size of the sampling window. Otherwise, the modulation can have a con-
siderable influence, even if the sampling window is much larger than the modulation
period.
8.6.1.4 Calculating the measurement time
Normally, the measurement time is calculated as follows:
MT = 2 * AC * APER + (2 * AC - 1) * 5
μ
s
With:
MT: overall measurement time
AC: average count
APER: aperture time
5 μs is the time for switching the chopper phase.
8.6.1.5 Configuring the result buffer
SENSe<Sensor>:[POWer:][AVG:]BUFFer:CLEar.................................................................57
SENSe<Sensor>:[POWer:][AVG:]BUFFer:COUNt?..............................................................57
SENSe<Sensor>:[POWer:][AVG:]BUFFer:SIZE...................................................................57
SENSe<Sensor>:[POWer:][AVG:]BUFFer:STATe.................................................................58
SENSe<Sensor>:[POWer:][AVG:]BUFFer:CLEar
Clears the contents of the result buffer.
Example:
BUFF:CLE
Usage: Event
SENSe<Sensor>:[POWer:][AVG:]BUFFer:COUNt?
Queries the number of results that are currently stored in the result buffer.
Example:
BUFF:COUN?
Usage: Query only
SENSe<Sensor>:[POWer:][AVG:]BUFFer:SIZE <count>
Sets the size of the result buffer.
You can enable the buffer using SENSe<Sensor>:[POWer:][AVG:]BUFFer:STATe
.
Parameters:
<count> Range: 1 to 8192
*RST: 1
Configuring the measurement modes