Remote control commands
R&S
®
NRPxxP
69User Manual 1179.5760.02 ─ 02
Parameters:
<lreference> Specified in percent of the pulse amplitude.
Range: 0.0 to 100.0
*RST: 10.0
Default unit: PCT
8.6.5.4 High-resolution measurements
SENSe<Sensor>:TRACe:ESAMpling:AUTO....................................................................... 69
SENSe<Sensor>:TRACe:MEASurement:TRANsition:ESAMpling:AUTO................................69
SENSe<Sensor>:TRACe:ESAMpling:AUTO <value>
Available in trace mode.
Enables or disables the automatic equivalent sampling that allows for high-resolution
measurements.
The automatic equivalent sampling requires a periodic measurement signal. The power
sensor superposes several successive measurements to improve the resolution. The
measurement speed decreases, because several successive measurements are nec-
essary.
Parameters:
<value> ON | OFF
*RST: ON
SENSe<Sensor>:TRACe:MEASurement:TRANsition:ESAMpling:AUTO <val>
Available in pulse analysis measurements.
Enables or disables the automatic equivalent sampling that allows for high-resolution
measurements.
The automatic equivalent sampling requires a periodic measurement signal. The power
sensor superposes several successive measurements to improve the resolution. The
measurement speed decreases, because several successive measurements are nec-
essary.
Parameters:
<val> OFF | ON
*RST: ON
8.6.5.5 Transferring measured pulse parameters
SENSe<Sensor>:TRACe:MEASurement:AUTO <mode>
Available in continuous average mode.
Enables or disables the automatic transfer of the measured pulse parameters after
each trace.