17 January, 2003 FIRMWARE UPDATE – CF CONFIGURATION
5-101
Service
Tables
6. After the firmware update is completed,
“Update done” or “Power Off On” is
displayed. Then, turn off the main switch
and remove the IC card.
7. If more firmware needs to be
downloaded, make sure that the main
switch is turned off and repeat steps 3
to 6.
8. When all firmware update is completed, remove the IC card while the machine
power is off and reinstall the IC card cover.
9. Turn the main switch on and confirm that the machine starts normally.
NOTE: Operation Panel Firmware Update
While the firmware of the operation panel
has been updating, the operation panel
cannot display anything (this is the normal
condition for firmware update processing
and completion).
The following message is displayed for 10
seconds after pressing the Up Date button,
then the message disappears and
firmware update starts.
You can check the firmware update processing and completion by watching the
Start key.
Condition of the Start key
When the above message disappears, the Start key starts blinking with red.
This shows you that the machine has started the firmware update, and it takes
about 5 minutes.
When the firmware update is completed, the Start key starts blinking with green
instead of red.
PCcard -> ROM Page01
Engine Card No.:1/1
Update done.
G080S514.WMF
Please do not turn the power off
until start key shows green blinking.
Starting program update of operation panel in few
seconds.
Install time is approximately 5 minutes.
G080S515.WMF