71Overview of the system
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
• Glucose test
• Emergency test
• OTS
• Proficiency test
• QC test
• Linearity test
• Insulin entry
Memory lockout is resolved by synchronizing the
instrument with the DMS to ensure it has sufficient free
memory.
User lockout
User lockout occurs when a user’s certificate expires and
they require recertification.
The instrument can be configured to display a user
certification expiration warning in advance of the
individual user certification expiration.
The lockout is displayed on the home screen when a
user requiring certification is logged in. The warning and
lockout is shown on the tab.
Only the observed test sequence (OTS) is available in this
lockout type.
The lockout is resolved by the user taking and passing an
OTS. As a result, the user’s name is added to an updated
user list which is distributed to all instruments. This
enables the user who was locked out to perform tests
again.
u User certification status (74)
u Performing an observed test sequence (OTS) (145)
2 Overview of the system