Trace Menu R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.30 E-6
DC MEAS 1V and DC MEAS 10V select the DC voltages fed to the input connectors DC
MEAS as measured quantities.
PAE... opens a dialog to select and configure the PAE measurement.
All DC and PAE measurements use the input connectors DC MEAS at the rear of the instrument. The
measurement results are real-valued and displayed as a function of sweep variable (frequency, internal
source power, time). The measurements can be performed in parallel to all RF measurements.
DC MEAS 1 V, DC MEAS 10 V
Select the DC voltages fed to the DC MEAS input connectors as measured quantities. The input
connectors are located at the rear panel of the instrument.
The two DC inputs cover different input voltage ranges; see data sheet or rear panel labelling.
The left input connector DC MEAS
1V provides the most accurate measurement for smaller
voltages but has a restricted input level range.
The right input connector DC MEAS
10V can be used for larger voltages.
Remote
control:
CALCulate<Ch>:PARameter:MEASure "<Trace_Name>", "DC+1V" |
"DC+10V"
[SENSe<Ch/Tr>:]FUNCtion[:ON] "...:VOLTage[:DC] DCIN1 | DCIN2"
Create new trace and select name and measurement parameter:
CALCulate<Ch>:PARameter:SDEFine "<Trace_Name>", "DC+1V" |
"DC+10V"
PAE...
Opens a dialog to select the Power Added Efficiency (PAE) of an active 2-port device as measured
quantity and to define the parameters for the PAE measurement.