ZVT Firmware Update
1300.0045.62 1.17 E-1
Firmware Update
Upgrade versions of the analyzer firmware are supplied as single setup files *.msi. To perform a
firmware update
1. Copy the setup file to any storage medium accessible from the analyzer. This may be the
internal hard disk, an external storage medium (USB memory stick, CD-ROM with external
drive) or a network connection (LAN, GPIB bus).
2. Start the installation using one of the following methods:
• If you use an external monitor or PC to display the GUI, access the Start menu of the operating system
by pressing the Windows key on the external keyboard or on the hardkey bar of the GUI and double-
click the setup file.
• Otherwise, use the remote control command SYSTem:FIRMware:UPDate '<setup_file_name>' to start
the installation.
3. Follow the instructions of the setup wizard.
Setup files can be stored and installed again. The default drive name of the internal hard disk is C:.
External storage devices are automatically mapped to the next free drive, i.e. D:, E: etc.
Attention!
The Calibration Unit (accessory R&S ZV-Z51) must be disconnected during a
firmware update.