ZVA/ZVB/ZVT Channel Menu
1145.1084.12 4.201 E-6
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
"<cal_name>", TOSM, <port_no>, <port_no>, <port_no>
Four Port P1 P2 P3 P4
Opens a submenu to select a four-port TOSM calibration at test ports PORT 1 to PORT 4:
Four-port calibrations can be performed automatically or manually.
Calibration Unit starts an automatic 4-port TOSM calibration. This function is enabled while the
Calibration Unit is connected.
Manual... calls up the start dialog of the calibration wizard to initiate a 4-port TOSM calibration.
Use Start Cal – Other... to select other 4-port calibration types.
Remote
control:
[SENSe<Ch>:]CORRection:COLLect:AUTO ' ', <port_no> (for matching
analyzer and cal unit ports)
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs (for arbitrary pairs of ports)
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
"<cal_name>", TOSM, <port_no>, <port_no>, <port_no>, <port_no>
Guided Calibration
The analyzer provides a calibration wizard for each calibration type. The guided calibration consists of
the following steps:
1. Select Physical Port Connectors and calibration kits at all calibrated ports.
2. Compile Calibrations: Select a calibration type and the physical ports to be calibrated (skipped for
predefined calibrations).
3. Measure Standards: Acquire measurement data for all standards required for the selected
calibration type.
4. Calculate the system error correction data (error terms) from the measurement data of the
standards and apply the result to the active channel.