Channel Menu R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.202 E-6
A successful calibration will supersede the previous calibration, discarding all previous system
error correction data. To keep older correction data you can transfer them into a Cal Pool using the
Calibration Manager.
The system error correction data determined in a calibration procedure are stored on the analyzer. You
can read these correction data using the remote control command
[SENSe<Ch>:]CORRection:CDATa. You can also replace the correction data of the analyzer by
your own correction data sets.
Select Connectors
The first dialog of the calibration wizard displays a table to select the connectors and calibration kits for
all calibrated physical ports.
The table contains the following rows:
Physical Port Number #
The ports (and therefore the number of table rows) are determined by the active calibration type
selected in the Start Cal submenu. For a user-defined calibration (Start Cal Other...) all
measurement ports of the analyzer are displayed regardless of their port configuration.