Trace Menu R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.34 E-6
• Show as selects the physical unit of the displayed trace. It is possible to display the measured
Voltage U or convert the wave quantity into an effective power according to P = V
2
/Re(Z
0
). Z
0
denotes the reference impedance of the source port (for wave quantities a
n
) or of the receive
port (for wave quantities b
n
). The reference impedances are defined in the Port Configuration
d
ialog.
• Source Port selects one of the available test ports of the analyzer as a source of the stimulus
signal.
Remote
control:
CALCulate<Ch>:PA
Rameter:MEASure "<Trace_Name>", "Pmtr1D1" |
...
Create new trace and select name and measurement parameter:
CALCulate<Ch>:PARameter:SDEFine "<Trace_Name>", "Pmtr1D1"
|...
Format
The Format submenu defines how the measured data is presented in the graphical display.
• dB Mag selects a Cartesian diagram with a logarithmic scale of the vertical axis to display the
magnitude of a complex measured quantity.
• Phase selects a Cartesian diagram with a linear vertical axis to display the phase of a complex
measured quantity in the range between –180 degrees and +180 degrees.
• Smith selects a Smith diagram to display an S-parameter or ratio.
• Polar selects a polar diagram to display an S-parameter or ratio.
• Delay calculates the group delay from an S-parameter or ratio and displays it in a Cartesian
diagram.
• Aperture sets a delay aperture for the delay calculation.
• SWR calculates the Standing Wave Ratio from the measured reflection S-parameters and
displays it in a Cartesian diagram.
• Lin Mag selects a Cartesian diagram with a linear scale of the vertical axis to display the
magnitude of the measured quantity.
• Real selects a Cartesian diagram to display the real part of a complex measured quantity.