Channel Menu R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.198 E-6
Start Cal
The Start Cal submenu selects the calibrated ports and the calibration type.
The calibration types depend on the number of test ports of the analyzer. For a four-port unit:
• One-Port P1 and One Port P2 open submenus to select a one-port calibration at test ports
PORT 1 and PORT 2, respectively.
• Two-Port P1 P2 opens a submenu to select a two-port calibration at test ports PORT 1 and
PORT 2.
• Three-Port P1 P2 P3 opens a submenu to initiate a three-port (TOSM) calibration at test ports
PORT 1, PORT 2 and PORT 3.
• Four-Port P1 .. P4 opens a submenu to initiate a four-port (TOSM) calibration at test ports
PORT 1, PORT 2, PORT 3 and PORT 4.
• Other... opens the calibration wizard to select a calibration type for an arbitrary number of ports.
Use Other... if you wish to perform several calibrations in a single measurement sequence and
with a minimum of effort. The Other... calibration wizard allows you to measure a particular standard
and reuse the results for different calibration types.
One-port to 4-port calibrations can be performed automatically using the Calibration Unit (accessory
R&S ZV-Z51). For calibrations with more than 4 ports use the calibration unit R&S ZV-Z58.
Remote
control:
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
"<cal_name>", REFLshort | FOPort | FRTRans | OPTPort | TOSM | TOM
| TRM | TRL | TNA, <port_no>[,<port_no>][,<port_no>][,<port_no>]