R&S
®
ZVA/ZVB/ZVT System Menu
1145.1084.12 4.261 E-6
GPIB Address
O
pens the numeric entry bar to define the analyzer's GPIB address.
The GPIB address must be in the range between 0 and 30.
Remote control:
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
Service Function...
Opens a dialog to access the service functions of the instrument. Service functions are password-
protected and should be used by a R&S service representative only. Refer to the service manual for
more information.
Remote control
(for channel bits):
SYSTem:PASSword[:CENable]
DIAGnostic:SERVice:FUNCtion
DIAGnostic:SERVice:SFUNCtion
Preset
Performs a preset of all instrument settings (i.e. all open setups) or of the active setup, depending on
the settings in the System Config dialog.
Scope of the preset and exceptions
A preset affects either the active setupor all open setups. It does not change the data related to global
resources (cal pool, cal kit data), the position of dialogs, the color scheme of the diagram areas, and the
directory and printer settings. All these properties can be reset in the Reset tab of the System Config
dialog.
If you activate Preset by mistake, you can use Undo in order to restore your previous instrument
settings.
Remote control:
SYSTem:PRESet
*RST
Undo
Reverses the last action, if possible. Otherwise, Undo! is disabled (grayed).
You can use Undo even after a Preset, in order to restore your own instrument settings.