EasyManua.ls Logo

Rohde & Schwarz R&S ZVA 8 - Show Limit Line

Rohde & Schwarz R&S ZVA 8
813 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S
®
ZVA/ZVB/ZVT Trace Menu
1145.1084.12 4.109 E-6
Show Limit Line displays or hides the limit line associated to the active trace.
Limit Check On activates or deactivates the limit check.
Fail Beep On activates or deactivates the acoustic signal indicating a limit excess.
Define Limit Line opens a dialog to define, save or recall limit lines.
TTL Out Pass 1/2 activates or deactivates a TTL output signal at the USER CONTROL
connector indicating whether the trace was passed or failed.
Horizontal Line displays or hides the horizontal line of the active trace and changes its position.
Limit lines are available for all Cartesian diagram types (Trace – Format). For polar diagrams, the
functions of the Lines submenu are grayed. The limit lines are hidden and the limit check is disabled
when a Cartesian trace format is replaced by a polar diagram.
Show Limit Line
Shows or hides the limit line associated to the active trace in a Cartesian diagram area. A checkmark
appears next to the menu item when the limit line is shown.
In the diagram, upper and lower limit lines can be displayed with different colors. Limit line segments
with disabled limit check (see Define Limit Line) can also be colored differently. The limit line colors are
defined in the Define User Color Scheme dialog (Display – Display Config. – Color Scheme...).