R&S
®
ZVA/ZVB/ZVT Trace Menu
1145.1084.12 4.71 E-6
Maximum Flatness
(Bohman)
46 dB 0 dB Maximum attenuation of responses outside the
gate span
Arbitrary Gate Shape
(Dolph-Chebychev)
User defined between
10 dB and 120 dB
0.071 dB Adjustment to individual needs; tradeoff between
sidelobe suppression and edge steepness
Remote
control:
CALCulate<Ch/Tr>:FILTer[:GATE]:TIME:STARt
CALCulate<Ch/Tr>:FILTer[:GATE]:TIME:STOP
CALCulate<Ch/Tr>:FILTer[:GATE]:TIME:CENTer
CALCulate<Ch/Tr>:FILTer[:GATE]:TIME:SPAN
CALCulate<Ch/Tr>:FILTer[:GATE]:TIME[:TYPE] BPASs | NOTCh
CALCulate<Ch/Tr>:FILTer[:GATE]:TIME:WINDow RECT | HAMMing | HANN |
BOHMan | DCHebyshev
CALCulate<Ch/Tr>:FILTer[:GATE]:TIME:DCHebyshev
Trace Statistics
Opens a submenu to evaluate and display statistical and phase information of the entire trace or of a
specific evaluation range and calculate the x-dB compression point.
Min/Max/Peak-Peak displays or hides the essential statistical parameters of the trace in the
selected evaluation range.
Mean/Std Dev displays or hides the arithmetic mean value and the standard deviation of the
trace in the selected evaluation range.
RMS displays or hides the RMS value of the trace in the selected evaluation range.
Phase Delay/El Length displays or hides the phase delay and the electrical length of the trace
in the selected evaluation range (Eval Range...).
Compression Point starts the x-dB compression point evaluation