Trace Menu R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.110 E-6
Display of the limit line and limit check are independent of each other: Hiding the limit line does
not switch off the limit check.
Remote control:
CALCulate<Ch/Tr>:LIMit:DISPlay[:STATe] ON | OFF
Limit Check On
Switches the limit check of the active trace on or off. A checkmark appears next to the
menu item when the limit check is enabled.
When the limit check is switched on, a PASS or FAIL message is displayed in the center of
the diagram. If the limit check fails at a measurement point, the two trace segments to
the left and right of the point can change their color. The Limit Fail Trace color is defined
in the Define User Color Scheme dialog (Display – Display Config. – Color Scheme...). An
acoustic signal (Fail Beep) and a TTL signal indicating pass or fail can be generated in
addition.