ZVA/ZVB/ZVT Channel Menu
1145.1084.12 4.233 E-6
Delay defines a delay time between the test ports and the reference plane.
Auto Length determines a length offset with the condition that the phase response of the
measured quantity is minimized.
The Zero Delay at Marker function overwrites the Offset parameters.
Remote control
(for Reset
Offsets):
[SENSe<Ch>:]CORRection:OFFSet<port_no>[:STATe] <numeric
_value>
Electrical Length
Opens a dialog to define the offset parameters for the different physical and logical test ports as
electrical lengths.
The electrical length is calculated as the product of the mechanical length and the square root of
the permittivity e
r
; see Offset parameters: Definition above. Open the Mechanical Length dialog to
change the permittivity.
Remote
control:
[SENSe<Ch>:]CORRection:EDELay<port_no>:ELENgth <numeric
_value>
Mechanical Length
Opens a dialog to define the offset parameters for the different physical and logical test ports as
mechanical lengths and permittivities.