EasyManua.ls Logo

Rohde & Schwarz R&S ZVA 8 - Standby and Ready State; Replacing Fuses

Rohde & Schwarz R&S ZVA 8
813 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S
®
ZVT Putting the Instrument into Operation
1300.0045.62 1.9 E-1
Standby and Ready State
The STANDBY toggle switch is located in the bottom left corner of the front panel.
After switching on the AC power, press the STANDBY key briefly to switch the analyzer from the
standby to ready state or vice versa.
In standby state, the right, amber LED is on. The standby power only supplies the power switch
circuits and the optional oven quartz (OCXO, 10 MHz reference oscillator, option ZVAB-B4, stock
no. 1164.1757.02). In this state it is safe to switch off the AC power and disconnect the instrument
from the power supply.
In ready state, the left, green LED is on. The analyzer is ready for operation. All modules are
power-supplied and the analyzer initiates its startup procedure.
Caution!
The instrument is still power-supplied while it is in standby mode.
Replacing Fuses
The DC inputs PORT BIAS at the rear panel are protected by four fuses IEC 127 - F 250 L
(250 mA flink).
To replace the fuses open the fuse holder by slightly turning the lid counterclockwise.
Replacement fuses are provided with the instrument.

Table of Contents

Related product manuals