System Menu R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.256 E-6
Add/Modify External Power Meter
S
elects and configures an external power meter.
The USB power meters supported at present are configured automatically, all input fields except
the optional Name field are disabled. A particular driver and serial number can be selected via
remote control.
Remote control:
SYSTem:COMMunicate:RDEVice:GENerator<gen_no>:DEFine
External Generators
Configures external generators with their connection type and device address.
Use of external generators, examples
An external generator can serve different purposes:
• Extended measurement functionality: Each external generator represents an additional source
port. External generators increase the number of RF input signals for the DUT. A typical
example is a mixer measurement with a 2-port analyzer, where an external generator provides
the LO input signal.