Trace Menu R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.94 E-6
Remote control:
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:EXECute MAXimum
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:RESult?
Min Search
Sets the active marker to the absolute minimum in the search range, i.e. to the smallest of all response
values. If a complex trace format (e.g. a polar diagram) is active, the marker is set to the measurement
point with the minimum magnitude.
By default the search range coincides with the sweep range. If the active trace contains no markers, a
marker Mkr 1 is created to indicate the search result.
Remote control:
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:EXECute MINimum
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:RESult?
Next Peak
Sets the active marker to the next maximum or minimum in the search range, depending on the current
search criterion.
If a Max Search or a Bandpass Search is active, then the marker is set to the next maximum.
The next maximum is the maximum with the largest response value that is below the current
marker response value.
If a Min Search or a Bandstop Search is active, then the marker is set to the next minimum. The
next minimum is the minimum with the smallest response value that is above the current
marker response value.
By default the search range coincides with the sweep range. If the active trace contains no markers, a
marker Mkr 1 is created to indicate the search result. Next Peak is disabled while a Target Search is
active.
Remote control:
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:EXECute NPEak
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:RESult?
Peak Search >
Sets the active marker to the next peak with higher stimulus value. If the active trace contains no
markers, a marker Mkr 1 created to indicate the search result. The peak criteria are defined via Define
Peak.
Remote control:
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:EXECute RPEak
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:RESult?
< Peak Search
Sets the active marker to the next peak with lower stimulus value. If the active trace contains no
markers, a marker Mkr 1 created to indicate the search result. The peak criteria are defined via Define
Peak.