ZVA/ZVB/ZVT Channel Menu
1145.1084.12 4.149 E-6
Triggered Meas. Sequence
Selects the measurement cycle or sequence of actions to be triggered.
Sweep means that each trigger event starts an entire sweep, according to the current sweep
configuration.
Sweep Segment means that each trigger event starts the next sweep segment within the
current sweep. If a sweep type other than Segmented Sweep is active, this option is equivalent
to Sweep.
Point means that each trigger event starts the measurement at the next sweep point.
Partial Measurement means that each trigger event starts the next partial measurement at the
current or at the next sweep point. If each measurement points requires only one partial
measurement , this option is equivalent to Point.
Relation to other sweep settings
Some sweep settings are logically incompatible with a particular selection of the Triggered Meas.
Sequence:
If a Time sweep is performed, the Triggered Meas. Sequence is always a sweep.
Alternate sweep mode only makes sense if the triggered measurement sequence comprises
more than one sweep point. If Point or Partial Measurement is selected, Alternate sweep mode
is switched off and vice versa.
The period of the trigger events should be adjusted to the triggered measurement sequence. If
the analyzer receives a trigger event while the last sequence is still running, the trigger event is
skipped and a message Trigger event missed is displayed.
Remote control:
TRIGger[:SEQuence]:LINK
Trigger Delay
Specifies a delay time between the trigger event and the start of the measurement.
The Delay time entered must be zero or positive, indicating that the trigger event precedes the start of
the measurement (Posttrigger).
Remote control:
TRIGger[:SEQuence]:HOLDoff
Single (All Chans)
Toggles between single sweep and continuous sweep mode.
In continuous mode (Single (All Chans) not selected), the analyzer measures continuously,