Channel Menu R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.150 E-6
repeating the current sweep.
• In single sweep mode, the measurement is stopped after the number of sweeps selected in the
Define Restart dialog.
Press Single to activate the single sweep mode and start a single sweep sequence. To start
further sweep sequences press Restart.
Remote control:
INITiate<Ch>:CONTinuous ON | OFF
Restart
Stops the current measurement and restarts a measurement sequence. In Single sweep mode a new
single sweep sequence is started.
Remote control:
INITiate<Ch>[:IMMediate]
Define Restart
Opens a dialog to configure the Single sweep mode.
• Number of Sweeps selects the number of sweeps to be measured in single sweep mode: either
one (Single Sweep) or a group of sweeps.
• Scope defines whether the active sweep mode (single or continuous) and the Number of
Sweeps are valid for the Active Channel only or for All Channels in the active setup. If All
Channels is selected, the number of sweeps in the single sweep sequence is equal to the
selected Number of Sweeps times the number of channels. The sequence starts with the first
sweep in channel no. 1.
Remote control:
[SENSe<Ch>:]SWEep:COUNt
INITiate<Ch>:IMMediate:SCOPe
Mode
The Mode menu defines the properties of the physical and logical (balanced) test ports, configures the
source signals and the receiver, and provides special measurement modes.