R&S
®
ZVA/ZVB/ZVT System Menu
1145.1084.12 4.257 E-6
Power calibration: An external generator can provide the reference signal for a source or
receiver calibration. A typical example is a receiver power calibration using a measured wave
b
1
.
Configured external generators can be selected as additional sources in the More Wave
Quantities,More Ratios, Port Configuration,and in the power calibration dialogs.
In order to connect external generators via GPIB, VXI-11, or other interface types, it is necessary
to install the Virtual Instrument Software Architecture (VISA) library. VISA provides the