R&S
®
ZVA/ZVB/ZVT Trace Menu
1145.1084.12 4.69 E-6
Grid type: Keep Sweep time Time domain
resolution
Unambiguous range
Stop freq. and no. of points
Freq. gap and no. of points
Stop freq. and approx. freq. gap
Remote
control:
CALCulate<Ch/Tr>:TRANsform:TIME:LPASs KFSTop | KDFRequency |
KSDFrequency
Time Gate
Switches the time gate defined via Define Time Gate on or off. An active time gate acts on the trace in
time domain as well as in frequency domain representation. Gat is displayed in the trace list while the
time gate is active.
The time gate is independent of the frequency window used to filter the trace prior to the time
domain transformation.
In time domain representation, you can use the time gate settings in order to
eliminate unwanted responses in your signal. After switching back to the frequency
domain, you will receive the frequency response of your DUT without the contribution of
the unwanted responses.
Remote control:
CALCulate<Ch/Tr>:FILTer[:GATE]:TIME:STATe <Boolean>