Channel Menu    R&S
®
ZVA/ZVB/ZVT 
1145.1084.12 4.210  E-6 
Calibration procedure  
The source power calibration requires an external power meter, to be connected via GPIB bus, USB or 
LAN interface. The power sensor can be connected directly at the reference plane or to any other point 
in the test setup where the signal power is known to be proportional to the power at the reference 
plane.   
The source power calibration involves several steps:  
5.  The analyzer performs a first calibration sweep at the source power that is likely to 
produce the target power (Cal Power) at the reference plane. A known attenuation or gain at 
the source port and in the signal path between the source port and the reference plane is taken 
into account. 
6.  The power measured at the reference plane is displayed in the calibration sweep 
diagram and compared to the Cal Power. The comparison serves as an input for the correction 
of the source power.  
7.  Steps 1 and 2 are repeated for a specified Number of Readings or after the deviation 
between the measured power at the reference plane and the Cal Power is below a specified 
Tolerance. 
8.  After the last calibration sweep, it is possible to perform an additional Verification 
Sweep in order to check the accuracy.  
After a power calibration one can expect the power at the reference plane to be within the range of 
uncertainty of the power meter. After a change of the sweep points or sweep range, the analyzer 
interpolates or extrapolates the calibration data.  
The Source Power Cal dialog provides the following control elements: 
• Calibrate – Source contains all analyzer ports or external generators which can be used as a 
source for the calibrated wave quantity. The list contains all analyzer ports Port 1 to  Port n.
Generators  (Gen 1, Gen 2, ...) must be configured explicitly in the System Configuration –