Channel Menu R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.212 E-6
The dialog provides the following control elements:
• Port n Cal Power (where n is the calibrated source port selected in the Source Power Cal
dialog) shows the target power of the calibration (Cal Power) and the Cal Power Offset. Both
parameters can be adjusted in the Port n Cal Power dialog, to be opened via Modify Cal
Power...
• The parameters in the Power Meter Readings panel provide stop criteria for the calibration
procedure: Maximum Number of Readings is a limit for the number of calibration sweeps.
Tolerance denotes the maximum deviation of the measured power from the Cal Power. The
calibration procedure is stopped if the Maximum Number of Readings is reached or if the
measured power is within the Tolerance.
• If Switch off All Other Sources... is checked, the power at all sources except the calibrated
source is switched off during the calibration. This is advisable especially if the power calibration
is intended for intermodulation measurements where different signals are combined.
Remote
control:
SOURce<Ch>:POWer<Pt>:CORRection:NREadings
SOURce<Ch>:POWer<Pt>:CORRection:COLLect:AVERage[:COUNt]
SOURce<Ch>:POWer<Pt>:CORRection:COLLect:AVERage:NTOLerance
SOURce<Ch>:POWer<Pt>:CORRection:OSOurces[:STATe]