Channel Menu    R&S
®
ZVA/ZVB/ZVT 
1145.1084.12 4.214  E-6 
Purpose of a receiver power calibration  
A receiver power calibration ensures that the power readings at a specified input port of the analyzer 
agree with the source power level calibrated at an arbitrary reference plane. Typically, the reference 
plane is at the input of the receiver so that the calibration eliminates frequency response errors in the 
calibrated receiver.  
A receiver power calibration generally improves the accuracy of power (wave quantity) measurements.   
Calibration procedure and calibration sweep 
The receiver power calibration is based on a wave with known power.  
• No external connection is needed for a receiver power calibration on the reference receiver 
(wave quantity a
n
with  Port n used as a source). Besides, the reference receiver is 
automatically calibrated together with any source power calibration of an analyzer port.  
• Calibration of a received wave b
n
involves an external test setup. The received wave to 
calibrate can be generated by the analyzer or an external generator.   
The receiver power calibration involves a single calibration sweep. The calibration sweep is performed 
with current channel settings but with a maximum IF bandwidth of 1 kHz. Smaller IF bandwidths are 
maintained during the calibration sweep; larger bandwidths are restored after the sweep.  
The analyzer measures the power at each sweep point, compares the result with the nominal power of 
the source, and compiles a correction table.  
An acoustic signal indicates the end of the calibration sweep. At the same time, a message next to the 
calibrated source indicates the status of the receiver power calibration. After a change of the sweep 
points or sweep range, the analyzer interpolates or extrapolates the calibration data.   
The Receiver Power Cal dialog provides the following control elements: 
• Wave Quantity to Calibrate provides two pull-down lists to select one of the transmitted waves 
a
n
or one of the received waves b
n
for the calibration. Each power calibration only applies to the 
selected wave quantity in the active measurement.  
• Receiver Power Correction Off can be selected/deselected as soon as power calibration data 
have been acquired for the active wave quantity. If the box is checked, the calibration data is 
kept in memory but not applied to the wave quantity. The Power Correction Off softkey disables 
the power correction for all receivers and sources.  
• Used Source contains all analyzer ports or external generators which can which can be used as 
a source for the calibrated wave quantity. The list contains all analyzer ports Port 1 to Port n.