ZVA/ZVB/ZVT   Channel Menu 
1145.1084.12 4.219  E-6 
The Calibration Kits dialog contains a series of buttons, each of them opening a dialog: 
• Import Kit imports a cal kit file containing the parameters for a new calibration kit. It is possible 
to load several kits, so the analyzer asks whether to set the imported calibration kit active.  
• Add Kit combines different standards to form a new calibration kit.   
• View/ Modify Kit adds or deletes standards in an imported or user-defined calibration kit and/or 
changes their parameters.  
• Delete Kit removes an imported or user-defined kit from the list of Calibration Kits in Use.
• Export Kit exports calibration kit data to a cal kit file.   
• Connector Types adds or deletes connector types.  
The  Calibration Kits in Use: table is used to select a calibration kit for each connector type. The pull-
down lists contain all kits for the individual connector types.   
Remote control:
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:SELect "<Ckit_Name>"
Cal Kit Parameter Types  
The analyzer uses three types of parameters to describe the calibration standards. The parameter type 
is the same for all standards in a kit and therefore annexed to the kit name:  
 
• Universal parameters (no label)  describe calibration kit models with highly standardized 
components so that the parameters are valid for all calibration kits of the model.  
• Typical parameters (labelled typical) approximately describe a calibration kit model. To 
correct for deviations between the standards, each kit of the model is individually measured and 
delivered with an additional, kit-specific parameter set. Therefore each typical parameter set 
<kit_name> typical is complemented by an additional parameter set <kit_name> containing 
optimized parameters for an individual kit.