1145.1084.12 6.293 E-6
SYSTem:ERRor:ALL?
Queries and at the same time deletes all entries in the error queue.
The entries consist of an error number and a short description of the error. Positive error numbers are
instrument-dependent. Negative error numbers are reserved by the SCPI standard; see section Error
Messages.
*RST value
–
SCPI, Command
Types
Confirmed, command or query
Example:
SYST:ERR:ALL?
Query all entries in the error queue. 0,"No error" is returned if the
error queue is empty.
SYSTem:FIRMware:UPDate '<file_name>'
Installs a firmware version stored in a NWA setup file (*.msi) on the analyzer. The installation is
automatic and does not require any further action to be taken.
'<file_name>'
String variable for the name and directory of a NWA setup file
(*.msi).
*RST value –
SCPI, Command
Types
Device-specific, no query
Example:
SYST:FIRM:UPD 'C:\Setup\ZVAB_01.10.msi’
Install the firmware version 01.10 from the setup file stored in the
Setup directoy of the analyzer's internal hard disk.
SYSTem:PASSword[:CENable] '<password>'
Sends a password to the analyzer enabling a class of service functions to function (Command
ENable). Service functions are activated with the commands of the DIAGnostic system and should be
used by a R&S service representative only. Refer to the service manual for more information.
'<password>'
Case-sensitive string variable. Sending an invalid password
generates error -221, (settings conflict).
*RST value – (neither the password nor the protection of service functions is
affected by *RST)
SCPI, Command
Types
Confirmed, no query
Example:
SYST:PASS "XXXX"
Enter password.