EasyManua.ls Logo

Rohde & Schwarz R&S ZVA 8 - Page 764

Rohde & Schwarz R&S ZVA 8
813 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TRIGger<Ch>... R&S
®
ZVA/ZVB/ZVT
1145.1084.12 6.302 E-6
MANual Trigger signal generated by pressing the Manual Trigger softkey.
*RST value IMMediate
SCPI,
Command
Types
Confirmed, command or query.
Example:
TRIG:SOUR MAN
Activate manual trigger mode. The analyzer starts the next
sweep when the Manual Trigger softkey is pressed.
TRIGger<Ch>[:SEQuence]:TIMer <numeric_value>
Sets the period of the internal periodic signal that can be used as a trigger source
(TRIGger<Ch>[:SEQuence]:SOURce TIMer).
<Ch>
Channel number. If unspecified the numeric suffix is set to 1.
<numeric_value>
Timer period
Range [def. unit] 1E-005 s to 13680 s [s]. UP and DOWN increment/decrement
the timer period in 10 ms steps.
*RST value 1 s
SCPI, Command
Types
Confirmed, command or query
Example:
TRIG:SOUR TIM
Activate internal timer as trigger source.
TRIG:TIM 0.1
Set a timer period of 100 ms.

Table of Contents