Spectrum Analysis
R&S
®
Scope Rider RTH
142User Manual 1326.1578.02 ─ 12
<Order of harmonic>, <limit value in % of fundamental>
●
Comments introduced by "#" are allowed.
●
If a limit is not provided for a harmonic, the harmonic is displayed and its results
are included in the export, but not in the limit check.
●
The fundamental is always displayed and need not be included in the list.
Example:
RTH,Version,1.0
Harmonics,Limit[%]
2,10.5
4,0.004
20,3.7
10
13,0.01
3,0.01
6.3.6.2 Export File Format for Harmonic Results
The harmonic results export file format is similar to a waveform export file; it is also
stored in CSV format. You can convert the comma-separated text to columns (see
Chapter 12.4.3.3, "Converting CSV to Excel Files", on page 276).
The harmonics results file contains some general header data and then a set of result
data for each harmonic. A maximum of 64 results can be provided, but the actual num-
ber of harmonics may be restricted (see "Limits, Loaded File" on page 139). Each set
of data contains the results for a specific harmonic in each input channel.
Table 6-7: Header data (converted to table)
Harmonic Results
[Description - not included in file]
Model RTH1004
SerialNumber xxxxx
Firmware Version '1.50'
ID CH1 CH2 CH3 CH4
Results for each channel
NbOfResults [n]
Number of results included in averaging and statis-
tics operations
RMS Unit
Unit used for RMS results
RMS
Current RMS of the signal power (
RMS Max
Maximum RMS of the signal power
RMS Min
Minimum RMS of the signal power
THDf/ THDr [%]
Current THD in relation to the RMS amplitude of the
fundamental frequency (THDf) or the total power of
the signal (THD
R
)
Harmonics Measurement (Option R&S RTH-K34)