Remote Control Commands
R&S
®
Scope Rider RTH
341User Manual 1326.1578.02 ─ 12
Parameters:
<UserY1>, <UserY2> Range: -100E+24 to 100E+24
Increment: 100E-6
*RST: 0
Default unit: V
CURSor:Y1AMplitude?
CURSor:Y2AMplitude?
Return the vertical values of the crossing points between the tracking cursors and the
source waveform.
Return values:
<ResultAmplitude1>
<ResultAmplitude2>
Range: -100E+24 to 100E+24
*RST: 0
Usage: Query only
CURSor:MEASurement<m>:RESult:ACTual?
Returns the result of the indicated cursor measurement.
Suffix:
<m>
.
1..2
Defines the measurement index. Two simultaneous cursor mea-
surements can be performed.
Return values:
<Result> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Usage: Query only
CURSor:MEASurement<m>:RESult:LIMit?
Indicates whether the measurement results are inside the measurement range, or out-
side (clipping).
Suffix:
<m>
.
1..2
Defines the measurement index. Two simultaneous cursor mea-
surements can be performed.
Return values:
<ResultLimit> INSide | OVERflow | UNDerflow | OVUNflow
*RST: INSide
Usage: Query only